QbD-Driven Development and Validation of a Novel RP-HPLC Method for Quantitative Determination of Coenzyme Q10 in Single and Combined Dosage Forms
1Teerthankar Mahaveer College of Pharmacy, Teerthankar Mahaveer University, Delhi Road, Moradabad, UP, India.
2Neelkanth Group of Institution, College of Pharmacy, Modipuram, Meerut, UP, India
Corresponding Author E-mail:vishalwardwaj@gmail.com
DOI : http://dx.doi.org/10.13005/ojc/410521
ABSTRACT:The objective of the study was to develop and validate an authentic RP-HPLC methods based on Quality by Design (QbD) for the quantitative analysis of Coenzyme Q10 (CoQ10) in both single and combined dosage forms with clomiphene citrate (CC). Risk assessment, Design of Experiments (DoE), and the creation of the Method Operable Design Region (MODR) were all part of a methodical QbD approach. Screening tests (n=18) improved flow rate, mobile phase composition, and column chemistry. The impact of flow rate and percentage CAN on Critical Quality Attributes (CQAs), including Retention Time (RT), Resolution (Rs), and Tailing Factor, was investigated using a DoE (32 factorial design). Under ideal chromatographic conditions, a C18 column (250 × 4.6 mm, 5 µm), 70% CAN, phosphate buffer (pH 3.0), and a flow rate of 1.0 mL/min with UV detection at 275 nm were used. The process was confirmed to satisfy ICH Q2 (R1) requirements. 86 minutes for CoQ10 and approximately 4.05 minutes for CC. Acceptance criteria were met by the system appropriateness metrics (Resolution >35, Tailing ~1.3, Theoretical Plates >30,000). Great linearity was demonstrated by the method (2.5–200 µm/mL; R2=0.9997 CoQ10, R2=0.9996 CC). Research on accuracy revealed recoveries at 80%, 100%, and 120% levels that fell between 98 and 102%. Precision was confirmed by intra- and inter-day %RSD being less than 2%. Minimal differences in system appropriateness with intentional changes in flow rate and %ACN were indicative of robustness. After 24 hours at room temperature and 48 hours in the refrigerator, the solutions held their stability. In comparison to previously published methods, its QbD–driven approach guarantees regulatory compliance and superior method performance.
KEYWORDS:Coenzyme Q10 (CoQ10); Clomiphene Citrate (CC); Method Validation; MODR; Quality by Design
Introduction
Coenzyme Q10 (CoQ10), sometimes referred to as ubiquinone, is a lipid-soluble benzoquinone that is essential for the transfer of electrons inside mitochondria and the generation of cellular energy 1. Because it improves mitochondrial bioenergetics and lowers oxidative stress, it has shown great therapeutic promise in cardiovascular disorders, neurological diseases, and reproductive health2.
![]() |
Figure 1: Chemical structure of Coenzyme Q10 and Clomiphene Citrate. |
In women with infertility, CoQ10 supplementation has been shown to improve ovarian response and oocyte quality3. One common selective estrogen receptor modulator (SERM) for ovulation induction in anovulatory infertility is clomiphene citrate (CC)4. Because of its synergistic effect, CoQ10 and CC combination therapy has drawn attention. CoQ10 improves antioxidant defense and mitochondrial function, which in turn improves the ovarian response to CC therapy5. 6CoQ10’s lipophilic nature, poor solubility and significant affinity for lipid membranes make quantitative quantification difficult, particularly when combined with CC7–9.
![]() |
Figure 2: QbD Tool and Life Cycle |
The method of choice for CoQ10 analysis has been reverse phase-high performance liquid chromatography (RP-HPLC); however, conventional approaches are not robust or optimized in a systematic manner10. To guarantee method robustness and regulatory compliance, the International Council for Harmonization (ICH) and regulatory bodies advise that analytical method development follow the Quality by Design (QbD) principles11. Systematic risk assessment, the determination of important parameters for the method (CMPs), and the creation of a Method Operable Design Region (MODR) via Design of Experiments (DoE) are all components of the QbD approach 12. QbD has been successfully used in recent research to construct RP-HPLC techniques for variety of pharmaceutical formulations, showing enhanced method comprehension, reproducibility, and control over important quality aspects 13,14. Nevertheless, no systematic report of a QbD-based analytical approach for the simultaneous measurement of CoQ10 and CC in combination dose forms has been published. In order to meet with ICH Q2 (R1) requirements, the current work set out to design and validate a QbD-guided reverse-phase performance liquid chromatography method for measuring CoQ10 quantitatively in both single and combined dosage forms with CC 15,16.
Materials and Methods
Chemicals and Reagents
Coenzyme Q10 (CoQ10) and Clomiphene Citrate (CC) working standards were obtained from certified pharmaceutical suppliers. All solvents, such as acetonitrile (ACN) and methanol, were of HPLC quality and were acquired from Merck (India). Trifluoroacetic acid (TFA), formic acid, ammonium acetate, and ammonium formate were all analytical-grade reagents that were utilized without additional purification. A Milli-Q filtration system (Millipore, USA) was used to create ultrapure water. The mobile phases were filtered through and degassed in an ultrasonic bath prior to use. 0.45 µm nylon filters.
Chromatographic conditions and instrumentation
A UV-visible detector and autosampler were included in the RP-HPLC system that was used for the chromatographic analysis. Data acquisition and processing were carried out using proprietary chromatography software. The optimized separation was achieved under the following conditions:
Table 1: Equipment and Chromatographic Factors
| Sr. No. | Parameter | Details |
| 1 | Column | Gemini C18, 250 × 4.6 mm, 5 µm |
| 2 | Mobile Phase | Phosphate Buffer (pH 3.0): ACN (gradient mode) |
| 3 | Flow Rate | 1.0 mL/min |
| 4 | Detection Wavelength | 275 nm |
| 6 | Injection Volume | 20µL |
| 6 | Temperature | Ambient (~25±2 °C) |
| 7 | Run Time | 20 minutes |
Analytical QbD Strategy (Development of Analytical Method by QbD Approach):
Risk Assessment
Ishikawa diagrams and Failure Mode Effect Analysis (FMEA) were used in a systematic risk assessment to determine Critical Method Parameters (CMPs) and Critical Quality Attributes (CQAs). 17, 18. The CMPs identified were the organic phase ratio (% acetonitrile), buffer pH, flow rate, and column chemistry, as these variables were anticipated to have a significant impact on the performance of the analytical method. The CQAs encompassed retention time (RT), resolution (Rs), tailing factor (T), and theoretical plates (N). These are critical metrics for evaluating the effectiveness, selectivity, and robustness of the method. The selection of these parameters was based on their direct influence on the method’s quality and reliability. They were further evaluated through a Quality by Design (QbD) approach to guarantee optimization and control of the method19,20.
Design of Experiments (DoE)
A full factorial design was used to evaluate the impact of two independent variables (Table 2) FFD (%ACN and flow rate) at three levels each on CQAs 18.
Table 2: full factorial design
| Sr. No. | Factor | Level-1 | Level 0 | Level+1 |
| 1 | %ACN | 60% | 70% | 80% |
| 2 | Flow rate (mL/min) | 0.9 | 1.0 | 1.1 |
Fifteen experimental runs were conducted. Selected results for the combined dosage form are shown in Table 3.
Table 3: Selected DoE Trials (Combined Dosage Form)
| Run | %ACN | Flow Rate (mL/min) | RT CoQ10(min) | RT CC (min) | Rs | Tailing |
| 1 | 60 | 0.9 | 14.52 | 4.25 | 37.8 | 1.29 |
| 8 | 68 | 1.0 | 12.86 | 4.05 | 35.6 | 1.32 |
| 11 | 70 | 1.0 | 12.70 | 4.04 | 35.3 | 1.31 |
| 15 | 72 | 1.1 | 12.10 | 4.00 | 34.2 | 1.38 |
The response surface analysis and desirability function facilitated the successful identification of the Method Operable Design Region (MODR). The ideal ranges for the key method parameters were identified as follows: the organic phase ratio (% acetonitrile) was optimal between 68% and 72%, a flow rate of 0.95 to 1.05 milliliters per minute, and the buffer pH between 2.8 and 3.2. The ranges delineate the design space where the method is anticipated to perform reliably and robustly.
Screening Trials
A total of 18 screening trials were performed to evaluate multiple columns and mobile phases under varying buffer and organic solvent conditions. C18 columns provided better resolution compared to CN or C8 phases. Final selection was based on chromatographic behaviour, peak symmetry, and resolution.
Table 4: Representative Column and Mobile Phase Screening Summary
| Run | Column Type | Mobile Phase | Observation |
| 4 | Gemini C18 | ACN:20 mM Ammonium Acetate (60:40), pH 3.0 | Good CC peak; CoQ10 split into peaks |
| 9 | Thermo C18 | ACN: 20 mM Ammonium Acetate | Poor separation |
| 18 | Gemini C18 | EA:ACN/MeOH: ACN +0.1% Formic Acid | Good chromatography resolution |
Sample and Standard Solution Preparation
Standard stock solutions (1000 µg/mL) of CoQ10 and CC were prepared in methanol and diluted with the mobile phase to obtain calibration standards in the range of 2.5-200 µg/mL. The required amount of single and mixed dose forms was dissolved in methanol to create sample solutions, which were then sonicated, filtered through 0.22 µm syringe filters, and then injected into the HPLC apparatus.
Method Validation
According to ICH Q2(R1) requirements, the method was validated using parameters like linearity, accuracy, precision, robustness, and stability.
Linearity: Excellent linearity between 2.5-200 µg/mL was demonstrated by the technique, with correlation values of: R2=0.9997 for CoQ10 and R2=0.999 for CC.
Accuracy and Precision
Table 5: Accuracy (Recovery Study)
| Sr. No. | Analyte | Level (%) | Recovery (%) | %RSD |
| 1 | CoQ10 (Single) | 100 | 100.8 | 0.9 |
| 2 | CoQ10 (Combined) | 120 | 101.8 | 0.9 |
| 3 | CC (Combined) | 80 | 99.0 | 1.3 |
Intra- and inter-day precision showed %RSD values <2% for all analytes.
Robustness
Robustness was verified by varying flow rate (±0.1 mL/min) and %ACN (±2%). All critical parameters (RT, Rs, and Tailing) remained within acceptable system suitability criteria.
Solution Stability
The results of stability testing showed that the solutions containing the analyte were stable for 24 hours at room temperature and for 48 hours when refrigerated (4–8 °C). The assay results in both instances revealed a deviation of under 1%, signifying that no considerable degradation occurred during the periods tested.
Result
QbD was used to design an RP-HPLC technique for CoQ10 in both single and mixed dose forms with CC. The plan comprised DoE optimization, screening studies, risk assessment, and the creation of the Method Operable Design Region (MODR).
Risk Factors Evaluation
Failure Mode Effects Analysis (FMEA) and an initial Ishikawa diagram were used in a systematic risk assessment to identify Critical Method Parameters (CMPs) and Critical Quality Attributes (CQAs). The identified CQAs included the retention times (RT) of Coenzyme Q10 (CoQ10) and Clomiphene Citrate (CC), the resolution (Rs) between the two analytes, the tailing factor (T), and the number of theoretical plates (N), which are all essential indicators of chromatographic performance. The CMPs determined to significantly influence these quality attributes were the composition of the mobile phase (percentage of organic phase, i.e., acetonitrile), the type and pH of the buffer, the flow rate, and the column chemistry. These parameters were further optimized to ensure robust method performance.
Optimization of DoE
To investigate the impact of Flow Rate and Organic Phase (% CAN) on CQAs, a 32 full factorial Design of Experiments (DoE) was undertaken.
Table 6: Factors and Levels
| Sr. No. | Factor | Level-1 | Level-2 | Level+1 |
| 1 | %ACN (Organic Phase) | 60% | 70% | 80% |
| 2 | Flow Rate (mL/min) | 0.9 | 1.0 | 1.1 |
DoE Results
DoE Trials for Combined Dosage Form:
Table 7: Layout of Proposed 15 Trials
| Run | %ACN | Flow Rate (mL/min) | RT(CoQ10) | RT(CC) | Resolution (Rs) | Tailing Factor |
| 1 | 60 | 0.9 | 14.52 | 4.25 | 37.8 | 1.29 |
| 2 | 60 | 1.0 | 14.00 | 4.20 | 37.1 | 1.30 |
| 3 | 60 | 1.1 | 13.52 | 4.15 | 36.8 | 1.28 |
| 4 | 65 | 0.9 | 13.60 | 4.12 | 36.5 | 1.31 |
| 5 | 65 | 1.0 | 13.20 | 4.10 | 36.0 | 1.30 |
| 6 | 65 | 1.1 | 12.90 | 4.08 | 35.9 | 1.31 |
| 7 | 68 | 0.9 | 13.00 | 4.10 | 35.7 | 1.32 |
| 8 | 68 | 1.0 | 12.86 | 4.08 | 35.6 | 1.32 |
| 9 | 68 | 1.1 | 12.60 | 4.06 | 35.1 | 1.33 |
| 10 | 70 | 0.9 | 12.80 | 4.05 | 35.4 | 1.32 |
| 11 | 70 | 1.0 | 12.70 | 4.04 | 35.3 | 1.31 |
| 12 | 70 | 1.1 | 12.40 | 4.02 | 35.0 | 1.33 |
| 13 | 72 | 0.9 | 12.60 | 4.03 | 34.8 | 1.35 |
| 14 | 72 | 1.0 | 12.30 | 4.01 | 34.5 | 1.36 |
| 15 | 72 | 1.1 | 12.10 | 4.00 | 34.2 | 1.38 |
Interpretation (Contrast with Single and Combined Dosage Form)
In the single dosage form with CoQ10, the retention time (RT) range was slightly wider, spanning from 12.6 to 14.8 minutes, because there was no interaction with Clomiphene Citrate (CC). At a flow rate of 1.0 mL/min, optimal chromatographic performance was realized at roughly 70% acetonitrile (ACN), resulting in a CoQ10 peak resolution of about 32.
In the combined formulation of CoQ10 and CC, the inclusion of CC slightly affected the retention time of CoQ10, requiring careful flow regulation to ensure a resolution exceeding 35. The best separation, with no increase in the tailing factor, was achieved at an ACN concentration of 68–70% and a flow rate of roughly 1.0 mL/min. The two dose forms’ Method Operable Design Regions (MODRs) were comparable, falling between 68 and 72% ACN and 0.95 and 1.05 mL/min. To guarantee consistent baseline separation and robustness of the method, narrower control limits were necessary for the combined dosage form.
Method Operable Design Region (MODR):
![]() |
Figure 3: Counter Plot-Retention Time (CoQ10) |
![]() |
Figure 4: Counter Plot-Resolution (CoQ10 vs CC) |
![]() |
Figure 5: Desirability Plot (MODR) |
Desirability function and contour plots defined the MODR (generated from DoE) as:
Design of Experiment Contour & Desirability Plots Generated:
The contour plot for retention time is depicted in Figure 3, which demonstrates that retention time diminishes as the percentage of acetonitrile (%ACN) and the flow rate both increases. As shown in Figure 4, the resolution contour plot indicates that the best resolution was reached at a flow rate of 1.0 mL/min and at around 70% ACN. The desirability plot shown in Figure 5 prominently features the Method Operable Design Region (MODR), which falls within the 68–72% ACN range and the flow rate of 0.95–1.05 mL/min. Furthermore, maintaining the buffer pH between 2.8 and 3.2 helped ensure consistent performance. The single and combined dosage forms both exhibited optimal chromatographic performance within this defined MODR.
Screening Studies
A series of trials were conducted to screen different columns, mobile phase systems, and pH conditions for both single and combined dosage forms.
Extensive chromatographic screening was undertaken to identify optimal conditions for the analysis of CoQ10 in both single and combined dosage forms. Various stationary phases, mobile phase compositions, and pH conditions were systematically evaluated. The Gemini, Phenomenex C18 column demonstrated consistent performance across multiple trials and emerged as a promising candidate for method development. Adjustments in the mobile phase composition, including the use of ammonium formate and ammonium acetate at different ratios and pH values, revealed key insights into peak behavior and selectivity. While some early runs exhibited poor peak resolution or incomplete separation, these trials provided critical direction for further optimization. The introduction of gradient systems using solvents such as ethyl acetate (EA), isopropanol (IPA), and acid modifiers (TFA and formic acid) contributed to improved chromatographic behavior. Notably, the combination of EA: ACN (50:50) and MeOH:ACN (80:20) with 0.1% formic acid on the Gemini C18 column (Run 18) produced well-resolved peaks and excellent chromatographic performance, confirming its suitability for subsequent method development.
Representative Chromatograms
![]() |
Figure 6: Optimized HPLC Chromatogram-Coenzyme Q10 (Single Dosage Form). |
![]() |
Figure 7: Optimized HPLC Chromatogram-Coenzyme Q10 + Clomiphene Citrate (Combined Dosage Form). |
The study concluded that C18 columns offered better peak resolution than CN columns. When acetonitrile (ACN) and ammonium acetate buffer (pH 3.0) were combined, it produced well-defined and acceptable peaks for both Coenzyme Q10 (CoQ10) and Clomiphene Citrate (CC). An optimal organic phase content of around 60–70% ACN was chosen for further method optimization, based on these findings.
Method Validation
Validation was conducted as per ICH Q2(R1) guidelines, covering linearity, accuracy, precision, robustness, and stability studies for both single and combined dosage forms.
Linearity
The technique demonstrated outstanding linearity for both analytes in the 2.5-200 µg/mL concentration range. Method reliability was shown by the correlation coefficients (R2), which were 0.9997 for CoQ10 and 0.9996 for CC method reliability.
![]() |
Figure 8: Linearity Plot- Coenzyme Q10 Click here to View Figure |
![]() |
Figure 9: Linearity Plot- Clomiphene Citrate |
![]() |
Figure 10: Combined Linearity Plot- Coenzyme Q10 and Clomiphene Citrate. |
Interpretation: Both analytes displayed a proportional increase in peak are with increasing concentrations, confirming method suitability for quantitative analysis across a broad range.
Accuracy and Precision
Accuracy
Accuracy was assessed by recovery studies at 80%, 100%, and 120% levels. All recoveries were within 98-102%, confirming method accuracy (Table 8).
Table 8: Accuracy
| Sr. No. | Analyte | Level (%) | Added (µg/mL) | Found (µg/mL) | Recovery (%) | %RSD |
| 1 | CoQ10 (Single) | 80 | 40 | 39.5 | 98.7 | 1.1 |
| 2 | CoQ10 (Single) | 100 | 50 | 50.4 | 100.8 | 0.9 |
| 3 | CoQ10 (Single) | 120 | 60 | 60.9 | 101.5 | 0.8 |
| 4 | CoQ10 (Combined) | 80 | 40 | 39.8 | 99.5 | 1.2 |
| 5 | CoQ10 (Combined) | 100 | 50 | 50.2 | 101.4 | 1.0 |
| 6 | CoQ10 (Combined) | 120 | 60 | 61.1 | 101.8 | 0.9 |
| 7 | CC (Combined) | 80 | 40 | 39.6 | 99.0 | 1.3 |
| 8 | CC (Combined) | 100 | 50 | 50.3 | 100.6 | 1.1 |
| 9 | CC (Combined) | 120 | 60 | 61.0 | 101.7 | 0.9 |
Precision
Precision was evaluated by intra-day and inter-day studies at two concentration levels (50 and 100 µg/mL). The %RSD was <2% for all runs, indicating reproducibility (Table 10).
Table 9: Precision
| Sr. No. | Analyte | Conc. (µg/mL) | Intra-day Mean ±SD | %RSD | Inter-day Mean ±SD | %RSD |
| 1 | CoQ10 (Single) | 50 | 50.2 ± 0.41 | 0.82 | 50.3 ± 0.55 | 1.09 |
| 2 | CoQ10 (Single) | 100 | 100.4 ± 0.79 | 0.78 | 100.5 ± 0.92 | 0.91 |
| 3 | CoQ10 (Combined) | 50 | 50.1 ± 0.44 | 0.88 | 50.4 ± 0.63 | 1.25 |
| 4 | CoQ10 (Combined) | 100 | 100.6 ± 0.91 | 0.9 | 100.8 ± 1.05 | 1.04 |
| 5 | CC (Combined) | 50 | 50.3 ± 0.46 | 0.91 | 50.5 ± 0.71 | 1.4 |
| 6 | CC (Combined) | 100 | 100.7 ± 0.95 | 0.94 | 101.0 ± 1.22 | 1.21 |
Robustness and Stability
Robustness was confirmed by the fact that purposeful changes in the mobile phase composition (±2%) and flow rate (±0.1 mL/min) had no discernible impact on system suitability metrics. For 24 hours at ambient temperature and 48 hours in a refrigerator, analytical solutions held their stability.
Table 10: Robustness Study (System Suitability under Deliberate Variations)
| Sr.no. | Paramete | Condition | RT (CoQ10) | RT (CC) | Resolution (Rs) | Tailing Factor | %RSD |
|
1 |
Flow Rate |
0.9 mL/min (-0.1) | 13.05 | 4.10 | 35.7 | 1.33 | 0.96 |
| 1.0 mL/min (Normal) | 12.86 | 4.05 | 35.6 | 1.32 | 0.89 | ||
| 1.1 mL/min (+0.1) | 12.56 | 3.98 | 35.1 | 1.34 | 1.02 | ||
|
2 |
Mobile Phase |
68.32 (-2%ACN) | 13.10 | 4.12 | 35.8 | 1.34 | 1.11 |
| 70.30 (Normal) | 12.86 | 4.05 | 35.6 | 1.32 | 0.89 | ||
| 72.28 (+2% ACN) | 12.60 | 4.00 | 34.9 | 1.36 | 1.20 |
Interpretation: All system suitability parameters remained within acceptable limits (Resolution >30, Tailing ~1.3, %RSD <2%, confirming method robustness.
Table 11: Stability Study
| Sr. No. | Storage Condition | Time | %Assay (CoQ10) | %Assay (CC) | %Deviation |
| 1 | Room Temperature | 0 h (Initial) | 100.0 | 100.0 | – |
| 24 h | 99.5 | 99.2 | <1% | ||
| 2 | Refrigerated (4-8 °C) | 24 h | 99.7 | 99.4 | <1% |
| 48 h | 99.3 | 99.0 | <1% |
Interpretation: No significant degradation observed; assay remained within ±2%, confirming solution stability.
Final Optimized Method:
Table 12: The final, improved chromatographic technique
| Sr. No. | Parameter | Optimized Condition |
| 1 | Column | C18 (250 × 4.6 mm, 5 µm) |
| 2 | Mobile Phase | Phosphate buffer (pH 3.0): ACN (Gradient) |
| 3 | Flow Rate | 1.0 mL/min |
| 4 | Wavelength | 275 nm |
| 5 | RT (CoQ10) | ~12.86 min |
| 6 | RT (CC) | ~4.05 min |
| 7 | Resolution | >35 |
| 8 | Tailing Factor | ~1.3 |
Discussion
For the quantitative study of CoQ10 in both single and mixed dose forms with CC, a QbD-based RP-HPLC technique was effectively created and validated.
Chromatographic Optimization, The retention times (RT) achieved with the optimized chromatographic method were approximately 12.86 minutes for Coenzyme Q10 (CoQ10) and around 4.05 minutes for Clomiphene Citrate (CC), under conditions of less than 70% acetonitrile (ACN) and a flow rate of 1.0 mL/min. In the case of the single-dosage formulation, CoQ10 showed a slightly wider RT range (14.8–12.6 minutes), probably because CC was not present. By contrast, the combined dosage form necessitated a narrower method operable design region (MODR) in order to keep the resolution (Rs) above 35. System Suitability: The resolution for the individual CoQ10 dosage was about 32, the tailing factor was roughly 1.3, and the theoretical plates were close to 28,000. The resolution was greater than 35, the tailing factor was approximately 1.3, and the theoretical plates exceeded 30,000 in the combined dosage form, indicating enhanced chromatographic performance with both analytes present21–24.
Linearity, the two dosage forms showed outstanding linearity, with correlation coefficients (R²) of 0.9997 for CoQ10 and 0.9996 for CC, within the concentration range of 2.5–200 µg/mL. Accuracy and Precision, the recovery values for all analytes fell between 98% and 102%, with relative standard deviation (%RSD) values under 2%. This confirms the method’s accuracy and reproducibility. Due to less matrix interference, the accuracy of the single CoQ10 dosage (100.8%) was slightly higher than that of the combined formulation (100.4%). Robustness and Stability, the system suitability remained largely unaffected by intentional changes to the method parameters, as evidenced by %RSD values staying under 1.5%. Solutions that were prepared demonstrated stability for a maximum of 24 hours at room temperature and for as long as 48 hours when refrigerated.
Optimized Method The final optimized chromatographic procedure was made using a C18 column that had dimensions of 250 × 4.6 mm and a particle size of 5 µm. The gradient mobile phase consisted of acetonitrile (ACN) and phosphate buffer (pH 3.0). A 275 nm wavelength was employed for detection, and the flow rate was maintained at 1.0 mL/min. Under these circumstances, the retention time for Coenzyme Q10 (CoQ10) was roughly 12.86 minutes, while the retention time for Compound CC was about 4.05 minutes. With a resolution of more than 35 and a tailing factor of about 1.3, the method showed excellent separation and strong peak symmetry.
Conclusion
The developed QbD-based RP-HPLC method is robust, accurate, precise, and stability indicating, suitable for routine quality control (QC) and stability testing of both single CoQ10 and combined CoQ10 + CC dosage forms. Its QbD-driven optimization ensures regulatory compliance and enhanced method understanding, making it superior to reported methods.
Acknowledgement
The authors sincerely thank Teerthankar Mahaveer College of Pharmacy, Teerthankar Mahaveer University, Delhi Road, Moradabad-244001, UP, India, for providing the facilities and support needed to conduct this research.
Funding Sources
The author(s) received no financial support for the research, authorship, and/or publication of this article.
Conflict of Interest
Regarding this paper, the authors have no financial or other conflicts of interest.
Ethical Approval
The present research work doesn’t involve any part work based on animal or animal parts or human subjects.
Data Availability
This research article contains all of the data that was produced and examined.
References:
- Bhagavan HN, Chopra RK. Plasma coenzyme Q10 response to oral ingestion of coenzyme Q10 formulations. Mitochondrion. 2007 Jun;7:S78–88.
CrossRef - Bhatti GK, Gupta A, Pahwa P, Khullar N, Singh S, Navik U, et al. Targeting mitochondrial bioenergetics as a promising therapeutic strategy in metabolic and neurodegenerative diseases. Biomed J. 2022 Oct;45(5):733–48.
CrossRef - Xu Y, Nisenblat V, Lu C, Li R, Qiao J, Zhen X, et al. Pretreatment with coenzyme Q10 improves ovarian response and embryo quality in low-prognosis young women with decreased ovarian reserve: a randomized controlled trial. Reproductive Biology and Endocrinology. 2018 Dec 27;16(1):29.
CrossRef - Hassan Z. Comparison between tamoxifen and clomiphene citrate for induction of ovulation and successful conception in polycystic ovarian syndrome. Int J Reprod Contracept Obstet Gynecol. 2015;1349–52.
CrossRef - Rodríguez-Varela C, Labarta E. Does Coenzyme Q10 Supplementation Improve Human Oocyte Quality? Int J Mol Sci. 2021 Sep 2;22(17):9541.
CrossRef - Xu Y, Nisenblat V, Lu C, Li R, Qiao J, Zhen X, et al. Pretreatment with coenzyme Q10 improves ovarian response and embryo quality in low-prognosis young women with decreased ovarian reserve: a randomized controlled trial. Reproductive Biology and Endocrinology. 2018 Dec 27;16(1):29.
CrossRef - Cui S, Luo K, Quan Y, Lim SW, Shin YJ, Lee KE, et al. Water-soluble coenzyme Q10 provides better protection than lipid-soluble coenzyme Q10 in a rat model of chronic tacrolimus nephropathy. Korean J Intern Med. 2021 Jul 1;36(4):949–61.
CrossRef - Giri S, Varshney KK, Srivastava R. Therapeutic Potential and Prospects of L-arginine in Various Diseases and its Clinical Intervention. Curr Drug ther. 2024 Aug;19(5):529–45.
CrossRef - Giri S, Chandra P. Therapeutic Potential of Natural Flavonoids: Pharmacological Targets, Signaling Pathways, Molecular Mechanisms, and Clinical Perspective on Parkinson’s Disease. Curr Drug ther. 2025 Feb;20(3):315–31.
CrossRef - Podar AS, Semeniuc CA, Ionescu SR, Socaciu MI, Fogarasi M, Fărcaș AC, et al. An Overview of Analytical Methods for Quantitative Determination of Coenzyme Q10 in Foods. Metabolites. 2023 Feb 14;13(2):272.
CrossRef - Muheem A, Shakeel F, Zafar S, Jahangir MA, Warsi MH, Jain GK, et al. Development and validation of stability indicating liquid chromatographic (RP-HPLC) method for estimation of ubidecarenone in bulk drug and formulations using quality by design (QBD) approach. Brazilian Journal of Pharmaceutical Sciences. 2018 Mar 5;53(4).
CrossRef - Dewi M, Pratama R, Arifka M, Chaerunisaa A. Quality By Design: Approach to Analytical Method Validation. Sciences of Pharmacy. 2022 Jun 27;1(1):38–46.
CrossRef - Goraksh TA, Gitaram VD. QbD-Based RP-HPLC Method Development and Validation for Quantitation of Rimegepant in Standard and Pharmaceutical Formulations. International Journal of Pharmaceutical Research and Applications. 2025 Feb;10(2):942–5.
CrossRef - Raju C, Bojan S, Chinthaginjala C, Dravidamani M, Dommaraju AK, Raghunathan K. Analytical Quality by Design Assisted Optimization of RP-HPLC Method for the Estimation of Palovarotene Drug Substance and Drug Product by Box–Behnken Design. Biosci Biotechnol Res Asia. 2025 Mar 25;22(1):313–24.
CrossRef - Mr. Vishal Kumar Pathak ProfPCProfBS. APPLICATION OF NOVEL QBD APPROACH IN ANALYTICAL METHOD DEVELOPMENT AND VALIDATION FOR THE QUANTITATIVE ANALYSIS OF COENZYME Q10 IN SINGLE AND COMBINED DOSAGE FORMS. International Journal of Emerging Technologies and Innovative Research (www.jetir.org | UGC and issn Approved). 2025 Mar;12(3):i394–404.
- Swartz Michael, Krull IS. Handbook of analytical validation. CRC Press; 2012. 206 p.
CrossRef - Bairagi A, Kothrukar R, Chikhale H, Kosanam S, Borse L. AQbD-novel strategy for analytical methods. Futur J Pharm Sci. 2024 Oct 4;10(1):138.
CrossRef - Subriadi AP, Najwa NF. The consistency analysis of failure mode and effect analysis (FMEA) in information technology risk assessment. Heliyon. 2020 Jan;6(1):e03161.
CrossRef - Garg NK, Sharma G, Singh B, Nirbhavane P, Katare OP. Quality by Design (QbD)-Based Development and Optimization of a Simple, Robust RP-HPLC Method for the Estimation of Methotrexate. J Liq Chromatogr Relat Technol. 2015 Oct 21;38(17):1629–37.
CrossRef - Saroj S, Jairaj V, Rathod R. Applying green analytical chemistry for development and validation of RP-HPLC stability indicating assay method for estimation of fenoverine in bulk and dosage form using quality by design approach. J Liq Chromatogr Relat Technol. 2017 Apr 21;40(7):340–52.
CrossRef - Lunetta S, Roman M. Determination of coenzyme Q10 content in raw materials and dietary supplements by high-performance liquid chromatography-UV: collaborative study. J AOAC Int. 2008;91(4):702–8.
CrossRef - Yadav K, Jatain I, Dubey KK, Nitharwal RG, Kaur I. Development of a thin layer chromatographic method for the determination of coenzyme Q 10 produced by Agrobacterium tumefaciens. Sep Sci Plus. 2023 Apr 29;6(4).
CrossRef - Podar AS, Semeniuc CA, Ionescu SR, Socaciu MI, Fogarasi M, Fărcaș AC, et al. An Overview of Analytical Methods for Quantitative Determination of Coenzyme Q10 in Foods. Metabolites. 2023 Feb 14;13(2):272.
CrossRef - Sayare AS, Undre PB, Ghode PD, Singh S V., Ghode SP. Development and Validation of RP-HPLC Method for the Estimation of Clomiphene Citrate in Pharmaceutical Dosage Form. Res J Pharm Technol. 2021 Jul 19;3483–8.
CrossRef
Accepted on: 04 Sep 2025
Second Review by: Dr. Manisha Rathore
Final Approval by: Dr. Murat Hatipoğlu