Quantitative Estimation of Negundoside in Nirgundi Taila by HPLC


Ajay Patel, U. K. Jain*, Nilesh Gupta and Ajay Raghuvanshi
Bhopal Institute of Technology and Science-Pharmacy, Bhojpur Road, Bhopal - 462 045 (India).

Download this article as:  PDF

ABSTRACT:

By optimizing the extraction, separation and analytical conditions, a reliable, rapid, simple and accurate liquid chromatography method with UV detection was developed for the quantitative determination of negundoside in an ayurvedic preparation Nirgundi taila. The estimation of negundoside was achieved on an reversed phase C18 column (250 mm X 4.6 mm ID, 5 μm particle size), with isocratic elution using a mixture of acetonitrile–methanol (70: 30, v/v) at a flow rate of 1 mL min-1 with UV detection at 254 nm, the sample injection volume was 10μL. The calibration curve was linear with correlation coefficients of 0.9996 for negundoside. The % Relative standard deviation (%RSD) value was less than 2% in the concentration range of 10–100 μg mL-1. Intra-day assay and inter-day assay precision of the analyte was less than 2%, and the average recovery rate obtained was in the range of 98–102% with %RSD below 2%. This method can provide a scientific and technical platform to the product manufacturers for setting up a quality control standard as well as to the public for quality and safety assurance of the proprietary ayurvedic formulations

KEYWORDS:

Column liquid chromatography; UV detection; Negundoside; Nirgundi Taila; Ayurvedic formulation