ISSN : 0970 - 020X, ONLINE ISSN : 2231-5039
     FacebookTwitterLinkedinMendeley

Abstract

On-Line Preconcentration by A Schiff’s Base Adsorbed on Surfactant Coated Activated Carbon and Determination of Nickel in Natural Water Samples by Flow Injection-Inductively Coupled Plasma Optical Emission Spectrometry (FI-ICP-OES)

Hossein Aref-Azar1* and Ali Moghimi2


Abstract:

An on-line nickel preconcentration and determination system implemented with inductively coupled plasma optical emission spectrometry (ICP-OES) associated to flow injection (FI) was studied. Trace amounts of nickel were preconcentrated by sorption on a conical minicolumn packed with activated carbon (AC) modified with a Schiff’s base 1,2-bis(salicylidene amino)ethane(SALET) at pH 5.0. The nickel was removed from the minicolumn with 20% nitric acid. An enrichment factor of 80-fold for a sample volume of 50 ml was obtained. The detection limit (DL) value for the preconcentration method proposed was 78 ng l-1. The precision for ten replicate determinations at the 0.5 μg l-1 Ni level was 2.5% relative standard deviation (R.S.D.), calculated from the peak heights obtained. The calibration graph preconcentration method for nickel was linear with a correlation coefficient of 0.9998 at levels near the detection limits (DL) up to at least 110 μg l-1. The method was successfully applied to the determination of nickel in natural water samples.

Keywords:

ICP-OES; Flow injection; On-line preconcentration; Nickel in water sample; 1,2-bis (salicyl idene amino)ethane(SALET)

[ View HTML Full Text]

Back to TOC